差分

このページの2つのバージョン間の差分を表示します。

この比較画面へのリンク

両方とも前のリビジョン前のリビジョン
次のリビジョン
前のリビジョン
research:memos:gr:gr_memo [2017/12/05 04:30] – [Tuning procedure] kobayashresearch:memos:gr:gr_memo [2022/05/21 21:28] (現在) kobayash
行 2: 行 2:
  
 ==== Preparation ==== ==== Preparation ====
 +
 +=== A/C ===
 +  * Call OGFA sfatt (ext.: 8890)
  
 === VDCs === === VDCs ===
行 7: 行 10:
   * Check if the preamps are turned on or not   * Check if the preamps are turned on or not
   * Check if the fans are turned on or not   * Check if the fans are turned on or not
 +  * The threshold of VDC HV supply is 40 uA for the proton beam at 300 MeV at E483 and E377+E422
 +  * The GR VDC threshold 4 uA --> 20 uA (sometimes 40 uA is used for 0 deg. exp.)
  
 === Viewers === === Viewers ===
行 92: 行 97:
       - Au target has some excited states at low energy. This looks a child of the beam?       - Au target has some excited states at low energy. This looks a child of the beam?
     - Changing the beam line setting takes time longer than 30 minutes (Nebiki-san & Furutaka-san)     - Changing the beam line setting takes time longer than 30 minutes (Nebiki-san & Furutaka-san)
-        - 50 minutes on Dec. 5, 2017 by Nebiki-san & Furutaka-san+      - 50 minutes on Dec. 5, 2017 by Nebiki-san & Furutaka-san
     - Optimization of beam energy resolution with the Au target (dispersion matching mode)     - Optimization of beam energy resolution with the Au target (dispersion matching mode)
       - This optimization should be done after the tuning for achromatic mode       - This optimization should be done after the tuning for achromatic mode
行 98: 行 103:
       - The goal is 15 keV (FWHM) without a target for proton faint beam at 295 MeV       - The goal is 15 keV (FWHM) without a target for proton faint beam at 295 MeV
       - Change the GR angle from 8.5 degree to 0 degree       - Change the GR angle from 8.5 degree to 0 degree
-      - SC FC is down+      - SC-FC is down
       - Faint beam       - Faint beam
       - The beam is centered (0 mm) in GX       - The beam is centered (0 mm) in GX
行 104: 行 109:
       - Y focus by changing the QM10 with Au target       - Y focus by changing the QM10 with Au target
       - Target Au --> blank       - Target Au --> blank
-      -  +      - QM10UDD optimization - First UD-M table which can realize good Y focus should be made 
 +        - UD (A) / M (A) / dE (keV) 
 +        - 25.50 / -41.20 / 31 keV 
 +        - 25.70 / -41.41 /  keV 
 +        - 25.90 / -41.41 /  keV 
 +        - 26.10 / -41.41 /  keV 
  
     - Q1-magnet focusing with the sieve slit and the carbon target     - Q1-magnet focusing with the sieve slit and the carbon target
     - Q1-magnet defocus with the sieve slit and the carbon target     - Q1-magnet defocus with the sieve slit and the carbon target
 +==== Restrictions ==== 
 +  * p 400 MeV - 10 nA (better less few nA) at SC-FC 
 +  * p 300 MeV - 20 nA at SC-FC 
 +  * 3He 420 MeV - 10 pnA at D1-FC (maybe too high) 
 +  * 
 ==== E404 ==== ==== E404 ====
   * Beam transportation   * Beam transportation
行 134: 行 149:
   * Dispersion matching tuning : 6 hours   * Dispersion matching tuning : 6 hours
   * In total 9 hours or something   * In total 9 hours or something
 +  * The Q magnet of the WS magnet should not be changed for the dispersion matching
 +  * The horizontal size of the beam spot at the target can not be narrow
 +
 +
 +
 +
 +
  
  
 ==== Resolutions ==== ==== Resolutions ====
   * Achromatic mode   * Achromatic mode
-    * p 300 MeV : 60-100 keV +    * p 295 MeV : 60-100 keV (FWHM) 
-    * 12C 1200 MeV : 240 - 400 keV+    * p 392 MeV : 80 keV (FWHM), best resolution 
 +    * 12C 1200 MeV : 240 - 400 keV (FWHM)
  
 ==== For 12C beam at 250 MeV (E404, Tsumura-san's exp) ==== ==== For 12C beam at 250 MeV (E404, Tsumura-san's exp) ====
行 157: 行 180:
   * Keep the GR and LAS chambers vacuum   * Keep the GR and LAS chambers vacuum
   * isobutane is used also by EN (but GR- and LAS-VDCs are the main users)   * isobutane is used also by EN (but GR- and LAS-VDCs are the main users)
 +  * Right-hand system for GR and LAS
 +  * Q1-FC angle coverage: 0 - ~20 deg.
 +
 +
  
-==== E483 ==== 
-  * VDC threshold 4 uA --> 20 uA (sometimes 40 uA is used for 0 deg. exp.) 
  
research/memos/gr/gr_memo.1512415820.txt.gz · 最終更新: 2017/12/05 04:30 by kobayash
CC Attribution-Share Alike 4.0 International
Driven by DokuWiki Recent changes RSS feed Valid CSS Valid XHTML 1.0